Accuracy: .995 over all ranges Pressure: Inlet - regulate to 20-50 psig, max 150 psig; vent - atmospheric not to exceed +5” water column Power: Specify 100/120 or 220/240 VAC Recovery Time: O2 Level Duration O2 Target Recovery on N2 9 ppm 1 minute 10 ppb 15 minutes 9 ppm 1 minute 1 ppb 60 minutes 1 ppm 5 minutes 1 ppb 30 minutes Response Time: 90% of final FS reading < 60 seconds Sample System: Manual pressure regulation, flow control and indicator. Pneumatic (minimum 80 psig air supply) valves control sample and span gas inlets, bypass and isolation of sensor, and, integral zero gas system. Electro-polished tubing with orbitally welded or VCR connections. Sensitivity: < 0.5% of FS range Sensor Model: GPR-13-2000UHP - requires no maintenance Sensor Life: 15 months at 25oC and 1 atm; average O2 < 10 ppm Signal Output: 4-20mA isolated and 0-1V Temp. Range: 5o to 45oC Warranty: 12 months analyzer; 12 months sensor Wetted Parts: 316L stainless steel Optional Equipment GPR-1600 UHP-B Delete integral zero gas system (see other side) GPR-1600 UHP-M Manual valves, delete zero gas system (see other side) Bezels for panel and 19” rack mounting; wall mount NEMA4X enclosure